Categories
YSTE-DH36 Hematology Analyzer
Item No.: YSTE-DH36
3-part differentiation, 21 parameters
Only 9ul sampling volume
Up to 60 samples per hour
50,000 sample results, including parameter, histogram, and patient information
Venous whole blood, anticoagulant peripheral whole blood and pre-diluted blood
Only 9ul sampling volume
Up to 60 samples per hour
50,000 sample results, including parameter, histogram, and patient information
Venous whole blood, anticoagulant peripheral whole blood and pre-diluted blood
Product parameters | |
Brand: | YSENMED |
MOQ: | 1 |
Package: | Wooden Package |
Ship From: | Guangzhou |
Lead Time: | 1-3 days |
Customization: | Yes |
Description
YSTE-DH36 Hematology Analyzer
Features:
3-part differentiation, 21 parameters
Only 9ul sampling volume
Up to 60 samples per hour
50,000 sample results, including parameter, histogram, and patient information
Venous whole blood, anticoagulant peripheral whole blood and pre-diluted blood
Technical Specifications:
Principles
Electrical impedance method (WBC/RBC/PLT) and cyanide free (HGB) method
Pulse baseline tracking technology, waveform discrimination technology, digital sheath flow technology and intelligent floating threshold technology
Dual-channel is used for counting
Parameters
WBC, Lym%, Mid#, Gran#, Lym%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,P-LCC , P-LCR, 3 histograms, alarm for abnormal erythrocytes, leukocyte, and platelet
Throughput
Up to 60 samples per hour
Sample Volume
Minimum volume 9 ul
Performance
Carryover Repeatability Linearity
WBC ≤0.5% ≤2.0% (7-15×109/L) 0.00-300×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.5×1012/L
HGB ≤0.5% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×109/L) 0.00-3000×109/L
Data Storage Capacity
Up to 50,000 results including parameter, histogram, and patient information
Communication
Four USB ports and Bi-directional LIS connection
Operating Environment
Temperature: 15oC-30oC
Humidity: 30-85%
Air Pressure: 70-106 kPa
Power Requirement
A.C. 100-240V ≤300VA 50/60Hz
Dimension and Weight
410mm(L) × 353mm(W) × 471mm(H)
Weight: 25kg
3-part differentiation, 21 parameters
Only 9ul sampling volume
Up to 60 samples per hour
50,000 sample results, including parameter, histogram, and patient information
Venous whole blood, anticoagulant peripheral whole blood and pre-diluted blood
Technical Specifications:
Principles
Electrical impedance method (WBC/RBC/PLT) and cyanide free (HGB) method
Pulse baseline tracking technology, waveform discrimination technology, digital sheath flow technology and intelligent floating threshold technology
Dual-channel is used for counting
Parameters
WBC, Lym%, Mid#, Gran#, Lym%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,P-LCC , P-LCR, 3 histograms, alarm for abnormal erythrocytes, leukocyte, and platelet
Throughput
Up to 60 samples per hour
Sample Volume
Minimum volume 9 ul
Performance
Carryover Repeatability Linearity
WBC ≤0.5% ≤2.0% (7-15×109/L) 0.00-300×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.5×1012/L
HGB ≤0.5% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×109/L) 0.00-3000×109/L
Data Storage Capacity
Up to 50,000 results including parameter, histogram, and patient information
Communication
Four USB ports and Bi-directional LIS connection
Operating Environment
Temperature: 15oC-30oC
Humidity: 30-85%
Air Pressure: 70-106 kPa
Power Requirement
A.C. 100-240V ≤300VA 50/60Hz
Dimension and Weight
410mm(L) × 353mm(W) × 471mm(H)
Weight: 25kg
YOUR MAY ALSO LIKE ...